Calibration of automatic potentiometric titrator
OBJECTIVE: To provide a procedure for calibration of automatic potentiometric titrator for moisture determination. SCOPE: Applicable to the automatic potentiometric titrator in Quality Control Department. Model : AT-38B+KF Manufacturer : Spectralab Sr. No. : S 302 RESPONSIBILITY: Chemist. PROCEDURE: Turn the instrument & printer on. Repeatability of factor determination. Go to titer mode. Use Disodium tartarate as standard. Take all precautions to avoid variations due to weighing & transfer of standard. Enter sample weight. Find out the factor. Take five readings. Calculate %RSD & average of five. %RSD of five readings should be less than 0.5. Calibration of Burette. Clean the ICB (Inter changeable burette). Fill water in reservoir. Fill the burette. Go to AUX functions & set Eject volume to 0.10ml. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Quit & then press Eject. Collect the volume in clean, dry, & previously weighed beaker. Weigh the beaker & find out the weight of ejected volume, further find out the volume. Volume = Weight in g / 0.99602 Volume should be between 0.0990ml to 0.1010 ml. Fill the burette. Go to AUX functions & set Eject volume to 1.0ml. Quit & then press Eject. Collect the volume in clean, dry, & previously weighed beaker. Weigh the beaker & find out the weight of ejected volume, further find out the volume. Volume = Weight in g / 0.99602 Volume should be between 0.99ml to 1.01 ml. Fill the burette. Go to AUX functions & set Eject volume to 10.0ml. Quit & then press Eject. Collect the volume in clean, dry, & previously weighed beaker. Weigh the beaker & find out the weight of ejected volume, further find out the volume. Volume = Weight in g / 0.99602 Volume should be between 9.98ml to 10.02 ml. Fill the burette. Go to AUX functions & set Eject volume to 20.0ml. Quit & then press Eject. Collect the volume in clean, dry, & pre weighed beaker. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Weigh the beaker & find out the weight of ejected volume, further find out the volume. Volume = Weight in g / 0.99602 Volume should be between 19.97ml to 20.03 ml. Precautions Clean the ICB with acetone after calibration of burette is over. FrequencyOnce in three months.REFERENCE:User’s guide for Automatic Potentiometric Titrator, Model: AT-38B+KFIndian Pharmacopoeia
|
No comments:
Post a Comment