Friday, December 30, 2011

Calibration of UV / Visible spectrophotometer


Calibration of UV / Visible spectrophotometer


OBJECTIVE:                                                                                                                      
To provide a procedure for calibration of UV / Visible spectrophotometer.
SCOPE: 
Applicable to the Spectrophotometer in Quality Control Department.
Model    : UV-1700, Pharmaspec.
Manufacturer : Shimadzu Corporation.
RESPONSIBILITY:
Chemist.
PROCEDURE:
Ensure that the connections of the instrument are proper.
Control of Absorbance
Potassium dichromate Solution :
Dry a quantity of potassium dichromate by heating to constant weight at 130°C. Weigh & transfer accurately a quantity not less than 57.0 mg & not more than 63.0 mg to 1000 ml volumetric flask.   Dissolve & dilute in sufficient 0.005M H2SO4 to produce 1000 ml.
Measure the absorbance of potassium dichromate solution at the wavelengths given below.
Calculate value of A (1% 1cm) for each wavelength.
A (1% 1cm) = Absorbance X 10000 / Weight of Potassium dichromate in mg.
Acceptance criteria


Wavelengths (nm)
A [1% 1cm]
Limit
235.0
124.5
122.9 - 126.2
257.0
144.0
142.8 - 145.7
313.0
48.6
47.0 - 50.3
350.0
106.6
105.6 - 108.2




Stringent limits are adopted from IP & BP for acceptance criteria
Resolution power
Record the spectrum of a 0.02% v/v solution of toluene in hexane in the range of 260 nm to 420 nm (before use check the hexane for transmittance, using water as a blank between 260nm to 420nm & use only if transmittance is not less than 97%).
Acceptance criteria:
The ratio of the absorbance at the maximum at about 269nm to that at the minimum at about 266nm is not less than 1.5.
Limit of Stray Light
Prepare a 1.2 % w/v solution of Potassium chloride in water.
Measure absorbance of the above solution at  198.0,199.0,200.0,201.0,202.0 nm using water as blank.
Acceptance criteria:
Absorbance is greater than 2.
Wavelength accuracy, Resolution & Baseline flatness (inbuilt test)
Attach printer directly to the instrument (instead of computer).
Go to Þ MODE press F3 key i.e. Maintenance.
Press ‘1’.
Press Start/Stop key.
After screen changes, ensure that nothing is kept in the optical path & press the Start/Stop key again.
Prints will come after all the three tests are over.
Acceptance criteria:
For wave length accuracy : at 656.1 ± 0.3 nm & at 486.0 ± 0.3 nm.
Resolution: 1.0 nm or less.
Baseline flatness: ± 0.002Abs.
Frequency

Once in a month.

Maintenance / Repairs
When the instrument does not comply with the requirement / tolerance range specified above, the instrument should be labeled “OUT OF CALIBRATION” and should be repaired / serviced.
After repair / maintenance, calibrate the instrument.

REFERENCE

Indian Pharmacopoeia
British Pharmacopoeia
Instruction manual of UV-1700.

CALIBRATION REPORT OF UV / VISIBLE SPECTROPHOTOMETER

Page 1 of 2
SOP Reference –
Date -

Instrument No. – EH/QCD/
Frequency
: Once in a month
Make –

Model –

1.
Control of Absorbance:
Potassium dichromate solution :
Gross Wt. :  ………………….                                 Balance : EH/QCD/
Tare Wt. :  ………………….
Net Wt.    : ………………….                                        ………..           ml with 0.005M sulphuric acid.
Dissolve and dilute 57.0 to 63.0 mg of potassium dichromate in sufficient 0.005M sulphuric acid to produce 1000.0 ml.

Wavelength (nm)
Absorbance
A (1%, 1cm)
Tolerance

235.0

 

 

122.9 to 126.2


257.0

 

 

142.8 to 145.7


313.0


47.0 to 50.3

350.0


105.6 to 108.2

Calculation :

             1%                    Absorbance  x 10000

          A1 cm    = -------------------------------------------------------

                          Weight of potassium dichromate in mg
Remark : Satisfactory / Not satisfactory.
Performed by :
Date :
Checked by :
Date :
2.
Resolution power :
0.02% v/v toluene in hexane : 
             ……….ml   ® 100.0 ml with hexane  / ………  ml   ® 100.0 ml with hexane
Absorbance at 269 nm = ……………
Absorbance at 266 nm = ……………
                          Absorbance at 269 nm
                Ratio = ---------------------------------- =    ---------------- =
                          Absorbance at 266 nm                             
  (Limit : Not less than 1.5)
Remark : Satisfactory / Not satisfactory.
Performed by :                                                                 Checked by:
Date :                                                                                Date           : 

CALIBRATION REPORT OF UV / VISIBLE SPECTROPHOTOMETER

Page 2 of 2
3.
Limit of stray light:
Potassium chloride solution
Gross Wt.                :  ………………….                                  Balance : EH/QCD/                           
Tare Wt. :  ………………….
Net Wt.    : ………………….                          ……………. ml with water.
Dissolve and dilute accurately 1.2 g potassium chloride  in sufficient water to produce 100.0 ml.

Sr. No.
Wavelength (nm)
Absorbance
Tolerance

1
198.0

Absorbance is greater than 2

2
199.0


3
200.0


4
201.0


5
202.0


Remark : Satisfactory / Not satisfactory.
Performed by :
Date :
Checked by :
Date :
4.
 Wavelength accuracy (In built test)
Sr. No.
Wavelength Standard (nm)
Wavelength Observed (nm)
Difference
Limit (nm)
1
656.1


+ 0.3
2
486.0


+ 0.3
Remark : Satisfactory / Not satisfactory.
Performed by :                                                                 Checked by:
Date :                                                                                Date           : 
5.
Resolution (In built test)
Sr. No.
Limit
Observation
1
1.0 nm or less
nm
Remark : Satisfactory / Not satisfactory.
Performed by :                                                                 Checked by:
Date :                                                                                Date           : 
6.
Baseline flatness (In built test)
Sr. No.
Limit
Observation
1
± 0.002 Absorbance

Remark : Satisfactory / Not satisfactory.
Performed by :                                                                 Checked by:
Date :                                                                                Date           : 

Conclusion : Instrument working is Satisfactory / Not satisfactory.

Next calibration due on :
















Performed by :
Date :
Checked by :
Date :

Calibration of automatic potentiometric titrator BY Dr.PIYUSH V. PATEL


Calibration of automatic potentiometric titrator


OBJECTIVE:
To provide a procedure for calibration of automatic potentiometric titrator for moisture determination.
SCOPE: 
Applicable to the automatic potentiometric titrator in Quality Control Department.
Model     : AT-38B+KF
Manufacturer : Spectralab
Sr. No.   : S 302
RESPONSIBILITY:
Chemist.
PROCEDURE:
Turn the instrument & printer on.
Repeatability of factor determination.
Go to titer mode.
Use Disodium tartarate as standard.
Take all precautions to avoid variations due to weighing & transfer of standard.
Enter sample weight.
Find out the factor.
Take five readings.
Calculate %RSD & average of five.
%RSD of five readings should be less than 0.5.
Calibration of Burette.
Clean the ICB (Inter changeable burette).
Fill water in reservoir.
Fill the burette.
Go to AUX functions & set Eject volume to 0.10ml.




















Quit & then press Eject.
Collect the volume in clean, dry, & previously weighed beaker.
Weigh the beaker & find out the weight of ejected volume, further find out the volume.
Volume = Weight in g / 0.99602
Volume should be between 0.0990ml to 0.1010 ml.
Fill the burette.
Go to AUX functions & set Eject volume to 1.0ml.
Quit & then press Eject.
Collect the volume in clean, dry, & previously weighed beaker.
Weigh the beaker & find out the weight of ejected volume, further find out the volume.
Volume = Weight in g / 0.99602
Volume should be between 0.99ml to 1.01 ml.
Fill the burette.
Go to AUX functions & set Eject volume to 10.0ml.
Quit & then press Eject.
Collect the volume in clean, dry, & previously weighed beaker.
Weigh the beaker & find out the weight of ejected volume, further find out the volume.
Volume = Weight in g / 0.99602
Volume should be between 9.98ml to 10.02 ml.
Fill the burette.
Go to AUX functions & set Eject volume to 20.0ml.
Quit & then press Eject.
Collect the volume in clean, dry, & pre weighed beaker.























Weigh the beaker & find out the weight of ejected volume, further find out the volume.
Volume = Weight in g / 0.99602
Volume should be between 19.97ml to 20.03 ml.
Precautions
Clean the ICB with acetone after calibration of burette is over.

Frequency

Once in three months.

REFERENCE:

User’s guide for Automatic Potentiometric Titrator, Model: AT-38B+KF

Indian Pharmacopoeia















A]
Repeatability of factor determination.

Sr. No.

Burette Reading (in ml.)

Factor
Average
% RSD
1




2


3


4


5


ACCEPTANCE CRITERIA
% RSD Not more than 0.5%.

B]

Calibration of Burette.

Sr. No.

Volume (ml.)

I

II

III

IV
Limit (ml.)

Empty Beaker Wt. (g.)

Wt. of Beaker + Water (g.)

II –I

III / 0.99602

1

0.1

 

 

 


0.0990 to 0.1010

2

1.0

 

 

 


0.99 to 1.01

3

10.0

 

 

 


9.98 to 10.02

4

20.0

 

 

 


19.97 to 20.03

 

CONCLUSION –
Calibration Complies / Does not comply as per Standard Calibration Procedure.

Next Calibration Due On -



Done By :
Date :
Checked By :
Date :