Calibration of automatic potentiometric titrator BY Dr.PIYUSH V. PATEL


Calibration of automatic potentiometric titrator


OBJECTIVE:
To provide a procedure for calibration of automatic potentiometric titrator for moisture determination.
SCOPE: 
Applicable to the automatic potentiometric titrator in Quality Control Department.
Model     : AT-38B+KF
Manufacturer : Spectralab
Sr. No.   : S 302
RESPONSIBILITY:
Chemist.
PROCEDURE:
Turn the instrument & printer on.
Repeatability of factor determination.
Go to titer mode.
Use Disodium tartarate as standard.
Take all precautions to avoid variations due to weighing & transfer of standard.
Enter sample weight.
Find out the factor.
Take five readings.
Calculate %RSD & average of five.
%RSD of five readings should be less than 0.5.
Calibration of Burette.
Clean the ICB (Inter changeable burette).
Fill water in reservoir.
Fill the burette.
Go to AUX functions & set Eject volume to 0.10ml.




















Quit & then press Eject.
Collect the volume in clean, dry, & previously weighed beaker.
Weigh the beaker & find out the weight of ejected volume, further find out the volume.
Volume = Weight in g / 0.99602
Volume should be between 0.0990ml to 0.1010 ml.
Fill the burette.
Go to AUX functions & set Eject volume to 1.0ml.
Quit & then press Eject.
Collect the volume in clean, dry, & previously weighed beaker.
Weigh the beaker & find out the weight of ejected volume, further find out the volume.
Volume = Weight in g / 0.99602
Volume should be between 0.99ml to 1.01 ml.
Fill the burette.
Go to AUX functions & set Eject volume to 10.0ml.
Quit & then press Eject.
Collect the volume in clean, dry, & previously weighed beaker.
Weigh the beaker & find out the weight of ejected volume, further find out the volume.
Volume = Weight in g / 0.99602
Volume should be between 9.98ml to 10.02 ml.
Fill the burette.
Go to AUX functions & set Eject volume to 20.0ml.
Quit & then press Eject.
Collect the volume in clean, dry, & pre weighed beaker.























Weigh the beaker & find out the weight of ejected volume, further find out the volume.
Volume = Weight in g / 0.99602
Volume should be between 19.97ml to 20.03 ml.
Precautions
Clean the ICB with acetone after calibration of burette is over.

Frequency

Once in three months.

REFERENCE:

User’s guide for Automatic Potentiometric Titrator, Model: AT-38B+KF

Indian Pharmacopoeia















A]
Repeatability of factor determination.

Sr. No.

Burette Reading (in ml.)

Factor
Average
% RSD
1




2


3


4


5


ACCEPTANCE CRITERIA
% RSD Not more than 0.5%.

B]

Calibration of Burette.

Sr. No.

Volume (ml.)

I

II

III

IV
Limit (ml.)

Empty Beaker Wt. (g.)

Wt. of Beaker + Water (g.)

II –I

III / 0.99602

1

0.1

 

 

 


0.0990 to 0.1010

2

1.0

 

 

 


0.99 to 1.01

3

10.0

 

 

 


9.98 to 10.02

4

20.0

 

 

 


19.97 to 20.03

 

CONCLUSION –
Calibration Complies / Does not comply as per Standard Calibration Procedure.

Next Calibration Due On -



Done By :
Date :
Checked By :